Bruker Avance 700 Reference and Temperature Calibration

Before Experiment


 

Probe:

5 mm TXI (1H/13C/15N)
5 mm QXI (1H/13C/15N/31P)

Calculate using:

Actual temperature:
Set temperature:
K

PPM
values:

Hppm :
Cppm :
Nppm :

 

 

 


Results:

Actual temperature: 300.00 K
Set temperature: 301.35 K
Hppm: 4.735
Cppm1: 56.000
Cppm2: 176.000
Nppm: 118.000
H DSSfreq: 700.1299820 MHz
Hfreq: 700.1332970 MHz
Cfreq1: 176.0572136 MHz
Cfreq2: 176.0783393 MHz
Nfreq: 70.9519249 MHz

 

 

After Experiment


 

Probe:

5 mm TXI (1H/13C/15N)
5 mm QXI (1H/13C/15N/31P)

Calculate using:

Actual temperature:
Set temperature:
K

Frequency
values:

Hfreq :
MHz
Cfreq :
MHz
Nfreq :
MHz

 

 


Results:

Actual temperature: 298.66 K
Set temperature: 300.00 K
Hppm: 4.752
Cppm1: 56.017
Cppm2: 176.018
Nppm: 118.017
H DSSfreq: 700.1299699 MHz
Hfreq: 700.1332970 MHz
Cfreq1: 176.0572136 MHz
Cfreq2: 176.0783393 MHz
Nfreq: 70.9519249 MHz